-30%
Beamex Accurate temperature dry block for industrial field use FB660R
18,549
26,499
Temperature block
Features
- We created the field temperature dry block (FB) for industrial field use.
- It is small and easy enough to carry around, provides fast calibration and also offers excellent accuracy.
- Combine it with the Beamex MC6 calibrator and Beamex calibration software to enable a fully automated and paperless calibration system.
| Hysteresis | ±0.01 °C |
|---|---|
| Immersion depth | 150 mm (5.9 in) |
| Insert OD dimensions | 24.4 mm (0.96 in) |
| Heating time: 15 min | 50 °C to 660 °C |
| Resolution | 0.01 °C / °F |
| Display | LCD, °C or °F user-selectable |
| Weight | 7.7 kg (17 lb) |
| Computer interface | RS-232 |
| Stander | Calibration Accredited calibration certificate provided |
| Environmental operating conditions | 0 °C to 50 °C, 0% to 90% RH (non-condensing) |
| Specifications valid in environmental conditions | 13 °C...33 °C |
| Temperature range at 23 °C | 50 °C to 660 °C (122 °F to 1220 °F) |
| Display accuracy | ±0.35 °C at 50 °C, ±0.35 °C at 420 °C, ±0.5 °C at 660 °C |
| Size | (H x W x D) 290 mm x 185 mm x 295 mm (11.4 x 7.3 x 11.6 in) |
| Stability | ±0.03 °C at 50 °C, ±0.05 °C at 420 °C, ±0.05 °C at 660 °C |
| Cooling time: 35 min | 660 °C to 50 °C, 25 min: 660 °C to 100 °C |
| Axial uniformity at 40 mm (1.6 in) | ±0.05 °C at 50 °C, ±0.35 °C at 420 °C, ±0.5 °C at 660 °C |
| Radial uniformity | ±0.02 °C at 50 °C, ±0.05 °C at 420 °C, ±0.10 °C at 660 °C |
| Loading effect (with a 6.35 mm reference probe and three 6.35 mm probes) | ±0.015 °C at 50 °C, ±0.025 °C at 420 °C, ±0.035 °C at 660 °C |
| Power requirements | 230 V (±10%), 50/60 Hz, 1800 W/100 V to 115 V (±10%), 50/60 Hz, 1400 W |
| Weight | 7.7 KG |
| ID | 1016362 |